Key specifications
- 12, 14, 16 bit resolution
- 50 MSa/s sampling
- 50 MHz bandwidth
- 128 Kpts memory per channel
- 100 ppm time base accuracy
![]() |
![]() |
![]() |
![]() |
![]() |
The Automotive Test Scope ATS5004D is a four channel automotive oscilloscope with differential inputs. Besides the normal features of a professional USB oscilloscope, the Automotive Test Scope ATS5004D has some specific features like SureConnect connection test and resistance measurement that make it an ideal instrument for automotive measurements.
Specifications
Acquisition system | |||||
---|---|---|---|---|---|
Number of input channels | 4 analog | ||||
Ch1, Ch2, Ch3, Ch4 | Isolated BNC, female | ||||
Type | Differential | ||||
Resolution | User selectable in software | ||||
Native | 12 bit | ||||
Enhanced | 14, 16 bit | ||||
Accuracy | 0.3 % of full scale ± 1 LSB | ||||
Noise | 135 µVrms (200 mV range, 12 bit, 50 MSa/s) 50 µVrms (200 mV range, 16 bit, 195 kSa/s) | ||||
Ranges (Full scale) | ±200 mV ±400 mV ±800 mV | ±2 V ±4 V ±8 V | ±20 V ±40 V ±80 V | ||
Coupling | AC/DC | ||||
Impedance | 2 MΩ / 40 pF | ||||
Maximum input voltage (in all ranges) | ±200 V (DC + AC peak < 10 kHz) with 1:10 attenuator ±300 V (DC + AC peak < 10 kHz) | ||||
Maximum Common Mode voltage | 200 mV to 800 mV ranges : 2 V 2 V to 8 V ranges : 20 V 20 V to 80 V ranges : 200 V | ||||
Common Mode Rejection Ratio | -48 dB | ||||
Bandwidth (-3dB) | DC to 50 MHz maximum | ||||
AC coupling cut off frequency (-3dB) | 1 Hz | ||||
Channel Isolation | 500 V | ||||
Channel Separation | -80 dB | ||||
SureConnect | |||||
Maximum voltage on connection | 200 V (DC + AC peak < 10 kHz) | ||||
Resistance measurement | |||||
Ranges (Full scale) | 100 Ω 200 Ω 500 Ω | 1 kΩ 2 kΩ 5 kΩ | 10 kΩ 20 kΩ 50 kΩ | 100 kΩ 200 kΩ 500 kΩ | 1 MΩ 2 MΩ |
Accuracy | 3% | ||||
Response time (to 95%) | < 5 ms | ||||
Maximum sampling rate | On all channels simultaneously | ||||
12 bit | 50 MSa/s | ||||
14 bit | 3.125 MSa/s | ||||
16 bit | 195.3 kSa/s | ||||
Maximum streaming rate | On all channels simultaneously | ||||
12 bit | 500 kSa/s | ||||
14 bit | 480.8 kSa/s | ||||
16 bit | 195.3 kSa/s | ||||
Sampling clock source | |||||
Internal | Quartz | ||||
Accuracy | ±0.01 % | ||||
Stability | ±100 ppm over -40 °C to +85 °C | ||||
Time base aging | ±5 ppm/year | ||||
External | On extension connector | ||||
Memory | 128 Kpts per channel | ||||
Trigger | Only available in scope/block mode, not in streaming mode | ||||
System | Digital, 2 levels | ||||
Source | CH1, CH2, CH3, CH4, OR, digital external | ||||
Trigger modes | Rising edge, Falling edge Inside window, Outside window | ||||
Level adjustment | 0 to 100 % of full scale | ||||
Hysteresis adjustment | 0 to 100 % of full scale | ||||
Resolution | 0.025 % (12 bits) | ||||
Pre trigger | 0 to 128 KiSamples, 1 sample resolution) | ||||
Post trigger | 0 to 128 KiSamples, 1 sample resolution) | ||||
Digital external trigger | |||||
Input | Extension connector | ||||
Range | 0 to 3.3 V (5 V max) | ||||
Coupling | DC | ||||
Interface | |||||
Interface | USB 2.0 High Speed (480 Mbit/s); (USB 1.1 Full Speed (12 Mbit/s) and USB 3.0 compatible) | ||||
Power Requirements | |||||
Power from USB port | 500 mA max (2.5 W max) | ||||
Power via external power input / extension connector | 1500 mA max (7.5 W max) | ||||
Minimum voltage | 4.5 VDC | ||||
Maximum voltage | 14 VDC | ||||
Physical | |||||
Instrument height | 25 mm (1 inch) | ||||
Instrument length | 170 mm (6.7 inch) | ||||
Instrument width | 140 mm (5.2 inch) | ||||
Cord length | 1.8 m (70 inch) | ||||
Weight | 460 g (16 ounce) | ||||
I/O connectors | |||||
Channel 1...4 | Isolated BNC, female | ||||
USB | fixed cable with USB 2.0 and USB 1.1 type A connector | ||||
Extension connector | D-sub 25 pins female | ||||
System Requirements | |||||
PC I/O connection | USB 2.0 High Speed (480 Mbit/s); (USB 1.1 Full Speed (12 Mbit/s) and USB 3.0 compatible) | ||||
Operating System | Windows 10 / 11, 64 bits | ||||
Operating Environment | |||||
Ambient temperature | 0 to 55 °C | ||||
Relative humidity | 10 to 90% non condensing | ||||
Storage Environment | |||||
Ambient temperature | -20 to 70 °C | ||||
Relative humidity | 5 to 95% non condensing | ||||
Certifications and Compliances | |||||
CE mark compliance | Yes | ||||
RoHS | Yes | ||||
Package | |||||
Convenient carry case | Carry case BB451 | ||||
Instrument | Automotive Test Scope ATS5004D | ||||
Probes | 4 x Measure lead TP-C1812B, BNC -> 4 mm banana plug | ||||
Accessoires | 4 x Differential attenuator TP-DA10 external power cable for second USB port | ||||
Software | Windows 10 / 11, 64 bits, via website | ||||
Drivers | Windows 10 / 11, 64 bits, via website | ||||
Manual | instrument manual and software user's manuals color printed on paper and digital, via website | ||||
Total package weight | Approx. 3 kg |