The Pico A3076 Series high-impedance active oscilloscope probe has minimal impact on the signal being probed with maximal signal transfer to the PicoScope 6000E Series. Its ergonomic design allows comfortable handheld use with a button to start and pause capturing in PicoScope 6.
The Intelligent Probe Interface powers the probe from the scope and facilitates automatic probe detection and unit scaling. The probe can set scope attenuation and automatically switch to 50 Ω input impedance.
With an input resistance of 1 MΩ and capacitance of 0.9 pF, this active probe offers high input impedance up to 750 MHz. These characteristics make this versatile probe ideal for many of your day-to-day measurements
Specifications
Probe bandwidth | DC to 750 MHz into 50 Ω load |
System bandwidth | DC to 750 MHz (nominal) system bandwidth with PicoScope 6000E Series 750 MHz oscilloscope |
Input dynamic range | ±5 V (DC + AC peak) |
DC gain accuracy | ±3% of signal |
DC gain accuracy (system) | ±4% of signal (nominal) with PicoScope 6000E Series oscilloscope |
DC offset accuracy | ±3 mV |
DC offset accuracy (system) | ±(1% of full scale + 4 mV) (nominal) with PicoScope 6000E Series oscilloscope. Offset accuracy can be improved by using the “zero offset” function in PicoScope 6. |
DC offset adjustment range | ±10 V |
DC offset control accuracy | ±1% of offset setting, additional to DC accuracy above |
Input characteristics | 1 MΩ || 0.9 pF |
Attenuation | 10:1 |
Oscilloscope input coupling | DC 50 Ω |
Dynamic performance (typical) | |
Harmonic distortion | –40 dBc (1%) with 250 MHz 1 V p-p sine |
Noise | 2.5 mV RMS (nominal) referred to probe input |
Bandwidth flatness | (+1 dB, –3 dB) from DC to full bandwidth |
Low-frequency flatness | < ±6% (or ±0.5 dB) from DC to 1 MHz |
Propagation delay | 5.4 ns (typical) |
Input voltage | |
Measurable voltage window | ±15 V (DC + AC peak) |
Maximum nondestructive input voltage | ±30 V (DC + AC peak) subject to frequency derating (derated as shown in A3000 Series Quick Start Guide). |
General | |
Probe detection | Automatic with any oscilloscope with Pico Intelligent Probe Interface (IPI) |
Probe button | Control start/stop capture in PicoScope 6 |
Operating temperature range | 0 to 50 °C |
Temperature range for quoted accuracy | 15 to 40 °C after 20-minute warmup (system performance specifications assume that both the probe and oscilloscope are within specified operating temperatures). |
Storage temperature range | –20 to +60 °C. |
Operating humidity range | 5% to 80% RH non-condensing |
Storage humidity range | 5% to 95% RH non-condensing |
Altitude range | Up to 2000 m |
Pollution degree | EN 61010 degree 2: “only nonconductive pollution occurs except that occasionally a temporary conductivity caused by condensation is expected” |
Dimensions (probe body) | 108 mm (L) x 16 mm (H) x 12 mm (W) |
Cable length | 1.2 m |
Input type | Sockets accepting 0.635 to 0.940 mm round or 0.64 mm square conductors. 7.12 mm pitch. |
Weight | 75 g |
Safety compliance | EN 61010-031:2015 |
EMC compliance | Tested to EN 61326-1:2013 and FCC Part 15 Subpart B |
Environmental compliance | RoHS, WEEE, REACH |