A3076 750 MHz active oscilloscope probe

Printer Friendly Tell a Friend
A3076 750 MHz active oscilloscope probe
Price: USD1,245.00

Qty:

  Description  

A3076 750 MHz active oscilloscope probe

Product Details

The Pico A3076 Series high-impedance active oscilloscope probe has minimal impact on the signal being probed with maximal signal transfer to the PicoScope 6000E Series. Its ergonomic design allows comfortable handheld use with a button to start and pause capturing in PicoScope 6.

The Intelligent Probe Interface powers the probe from the scope and facilitates automatic probe detection and unit scaling. The probe can set scope attenuation and automatically switch to 50 Ω input impedance.

With an input resistance of 1 MΩ and capacitance of 0.9 pF, this active probe offers high input impedance up to 750 MHz. These characteristics make this versatile probe ideal for many of your day-to-day measurements

Specifications

Probe bandwidth DC to 750 MHz into 50 Ω load
System bandwidth DC to 750 MHz (nominal) system bandwidth with PicoScope 6000E Series 750 MHz oscilloscope
Input dynamic range ±5 V (DC + AC peak)
DC gain accuracy ±3% of signal
DC gain accuracy (system) ±4% of signal (nominal) with PicoScope 6000E Series oscilloscope
DC offset accuracy ±3 mV
DC offset accuracy (system) ±(1% of full scale + 4 mV) (nominal) with PicoScope 6000E Series oscilloscope. Offset accuracy can be improved by using the “zero offset” function in PicoScope 6.
DC offset adjustment range ±10 V
DC offset control accuracy ±1% of offset setting, additional to DC accuracy above
Input characteristics 1 MΩ || 0.9 pF
Attenuation 10:1
Oscilloscope input coupling DC 50 Ω
Dynamic performance (typical)
Harmonic distortion –40 dBc (1%) with 250 MHz 1 V p-p sine
Noise 2.5 mV RMS (nominal) referred to probe input
Bandwidth flatness (+1 dB, –3 dB) from DC to full bandwidth
Low-frequency flatness < ±6% (or ±0.5 dB) from DC to 1 MHz
Propagation delay 5.4 ns (typical)
Input voltage
Measurable voltage window ±15 V (DC + AC peak)
Maximum nondestructive input voltage ±30 V (DC + AC peak) subject to frequency derating (derated as shown in A3000 Series Quick Start Guide).
General
Probe detection Automatic with any oscilloscope with Pico Intelligent Probe Interface (IPI)
Probe button Control start/stop capture in PicoScope 6
Operating temperature range 0 to 50 °C
Temperature range for quoted accuracy 15 to 40 °C after 20-minute warmup (system performance specifications assume that both the probe and oscilloscope are within specified operating temperatures).
Storage temperature range –20 to +60 °C.
Operating humidity range 5% to 80% RH non-condensing
Storage humidity range 5% to 95% RH non-condensing
Altitude range Up to 2000 m
Pollution degree EN 61010 degree 2: “only nonconductive pollution occurs except that occasionally a temporary conductivity caused by condensation is expected”
Dimensions (probe body) 108 mm (L) x 16 mm (H) x 12 mm (W)
Cable length 1.2 m
Input type Sockets accepting 0.635 to 0.940 mm round or 0.64 mm square conductors. 7.12 mm pitch.
Weight 75 g
Safety compliance EN 61010-031:2015
EMC compliance Tested to EN 61326-1:2013 and FCC Part 15 Subpart B
Environmental compliance RoHS, WEEE, REACH

 Data Sheet        User's Guide


Note: Specifications valid over ambient temperature range 15 to 40°C after 20-minute warmup.