Highlights:
  • 25 GHz sampling oscilloscope
  • 4 channels
  • Add External PG900 TDR/TDT Source

  Description  

9341-25

Product Details

Sampling Oscilloscopes to 25 GHz with TDR/TDT and Optical models
15 to 25 GHz electrical, 9.5 GHz optical, TDR/TDT, 2-channel and 4-channel, compact, portable, USB instruments.
These units occupy very little space on your workbench and are small enough to carry with your laptop for on-site testing, but that’s not all. Instead of using remote probe heads attached to a large bench-top unit, you can position the scope right next to the device under test. Now all that lies between your scope and the DUT is a short, low-loss coaxial cable. Everything you need is built into the oscilloscope, with no expensive hardware or software add-ons to worry about.

Key specifications

  • 15 TS/s (64 fs) sequential sampling
  • Up to 15 GHz prescaled, 2.5 GHz direct trigger and 11.3 Gb/s clock recovery
  • Industry-leading 16-bit 1 MS/s ADC and 60 dB dynamic range
  • Eye and mask testing to 16 Gb/s with up to 223–1 pattern lock
  • Intuitive, touch-compatible Windows user interface
  • Comprehensive built-in measurements, histogramming and editable data mask library
  • Integrated, differential, deskewable TDR/TDT step generator

Typical applications

  • Telecom and radar test, service and manufacturing
  • Optical fiber, transceiver and laser testing
  • RF, microwave and gigabit digital system measurements
  • Radar bands I, G, P, L, S, C, X, Ku
  • Precision timing and phase analysis
  • Digital system design and characterization
  • Eye diagram, mask and limits test to 10 Gb/s
  • Ethernet, HDMI 1, HDMI 2, PCI, SATA, USB 2.0, USB 3.0
  • TDR/TDT analysis of cables, connectors, backplanes, PCBs and networks
  • Optical fiber, transceiver and laser test
  • Semiconductor characterization

Specifications

General
Temperature range, operating +5 °C to +35 °C
Temperature range for stated accuracy Within 2 °C of last autocalibration
Temperature range, storage −20 °C to +50 °C
Calibration validity period 1 year
Power supply voltage +12 V DC ± 5%
Power supply current 1.7 A max.
Mains adaptor Universal adaptor supplied
PC connection USB 2.0 (compatible with USB 3.0)
LAN connection 10/100 Mbit/s
PC requirements Microsoft Windows XP (SP2 or SP3), Vista, 7, 8 or 10.
32‑bit or 64‑bit versions.
Dimensions 170 mm x 285 mm x 40 mm (W x D x H)
Weight 1.3 kg max.
Compliance FCC (EMC), CE (EMC and LVD)
Warranty 5 years