M3510A Digital Multimeter Print this pagePrint this page
M3510A Digital Multimeter
Price:

Model: M3510ARC (+ USD649.00); M3510AGC (+ USD700.00); M3510A (+ USD625.00)
Accessories: M3500-opt11 (+ USD45.00); M3500-opt10 (+ USD75.00); M3500-opt08 (+ USD99.00); M3500-opt07 (+ USD110.00)
Highlights:
  • 6 1/2 Digits High-Performance DMM
  • High Speed: 50K RDGS/S
  • Dual Measurements & Displays
  • Temperature Measurements
  • Multi-Measurement & Math Functions
  • True RMS
  • Easy-to-Use MENU
  • Built-in USB Interface & USBTMC Conformance
  • Quality Assurance
  • Free Application Software

Description

M3510A Digital Multimeter (Also Available: Model M3500A)

Product Details

The M3510A 6 1/2 Digit Multimeter is the newest generation of DMM from Picotest. This new DMM offers compact size, light weight, fast measurement speed & throughput, expanded measurement functions, systematization capability and a highly usable interface including a standard USB interface and a GPIB/RS-232 interface. The M3510A DMM is suitable for both production and research/development use. Like all Picotest products, it’s backed by our 30 day refund policy.

What's included with all models
CD(User manual & software applications), power cord, test leads, and USB cable.
Three (3) Year Manufacturer's Warranty

Model Description
M3510A 6.5 Digits High Speed Digital Multimeter
M3510AGC 6.5 Digits High Speed Digital Multimeter with GPIB
M3510ARC 6.5 Digits High Speed Digital Multimeter with RS232
Accessories Description
M3500-opt07 Kelvin Probe
M3500-opt08 4-Wire Test Leads
M3500-opt10 Shorting Plug
M3500-opt11 K Type Thermocouple Probe

Detailed Specifications
6 1/2 Digit High-Performance DMM
The M3510A 6 1/2 Digit Multimeter is the newest generation of DMM from Picotest. This new DMM offers a compact size, light weight, fast measurement speed & throughput, expanded measurement functions, systematization capability and highly usable interface including a standard USB interface and a GPIB/RS-232 interface. The M3510A DMM is suitable for both production and research/development use.

50000 Readings/Sec Sampling Rate
The M3510A adopts the latest A/D technology and achieves 50,000 readings per second at 4 1/2 digits. When sending data to your computer, the M3510A can achieve high precision and stability compared to other DMMs with slow data transmission rate. With the M3510A your work efficiency will be greatly improved, and the operational cost will be greatly reduced.

Dual Measurements & Displays
The dual measurement display feature of M3510A offers two measurement results displayed at the same time and it is easy to set up and configure. For example, when measuring DCV, you may also conduct Thermocouple measurement on M3510A. This allows you to observe temperature change while making other measurements. The M3510A can function as a multimeter and a thermometer in one.

Temperature Measurements
The M3510A DMM supports the RTD temperature measurement and supports the standards such as ITS-90, IEC751 and Callendar-Van Dusen. The RTD function is recommended for more precise measurement. The M3510A also provides the Thermocouple (TC) function. It has a built-in cold junction compensator for improved TC accuracy. Without additional plug-in's and reference temperature settings, you can measure temperature directly via Thermocouples of K, J, R, S, T, E, N, C and B types.

Multi-Measurement & Math Functions
The M3510A DMM offers the capacitance measurement and dual measurement display, in addition to standard measurements, such as DCV, DCI, ACV, ACI, 2WΩ, 4WΩ, Frequency, Period, Thermocouple & RTD, and the math functions, such as Limits, Ratio, MX+B, %, dBm, dB, Min/Max & Null. Measurement ranges have also been expanded. For instance, DC and AC Current Ranges now go up to 10 A.

True RMS
The M3510A DMM adopts an analog conversion technique to provide true RMS value for ACV/ACI measurements regardless of the waveform shape. Only the “heating value” of the AC components of the input waveform is measured. For non-symmetrical waveforms, such as pulse trains, the DC components will be rejected by the RMS measurement.

Simple to Use
Configuring the M3510A is straightforward. Unlike other DMMs with complicated operation modes in the configuration menu, M3510A offers easy-to-use SHIFT, CONFIG, ENTER buttons, arrow keys and corresponding soft keys at the lower part of the display across all the measurement configurations. Using equipments from Picotest just gets easier than ever.

Built-in USB Interface & USBTMC Conformance
A high-speed USB interface is built into the M3510A. A GPIB/RS-232 interface is also included. The included USB interface conforms to the USBTMC protocol. The M3510A can be operated through USBTMC compliant software. So get the M3510A now to replace your old DMMs while keep your existing USBTMC compliant software.

Reliability
The M3510A DMM is of high reliability and complies with CE requirements, from well-packed cartons with the shock absorbing bumpers to careful component selection, circuit protection design, rigorous environmental tests and ISO9001 production. Defective products can be returned with free repair/calibration with one-year warranty.

A Sample Application (Current vs. Temperature)
The M3510A DMM is shipped with the PT-TOOL & PT-LINK software. With the software, you can quickly establish an error-free connection between your PC and the equipment. The M3510A conforms to Standard Commands for Programmable Instrumentation (SCPI). PT-TOOL is a virtual oscilloscope software, which can monitor the DUT with scope display style. PT-LINK is used to collect measurement data in Microsoft Excel or Word and analyze it with charts. In addition, engineers are allowed to compile our Labview Driver for specific applications.

Dimension Information
If problems occur while operating the M3510A, it is recommended to do the self-test first. If you need to perform basic adjustments for Zero & Gain, please follow the instructions in the M3510A Service Manual. A calibrator with at least 6 1/2-digit precision and a shorting plug are required. For a complete calibration with factory reports, contact your local service representative to return your DMM to PICOTEST.

Accessories
CD (user manual and software applications), power cord, test leads, and USB cable.
GPIB Interface Card
3 Year Warranty
The following options are available:

  • M3500-opt04: GPIB Card
  • M3500-opt06: RS-232 Card
  • M3500-opt07: Kelvin Probe
  • M3500-opt08: 4-Wire Test Leads
  • M3500-opt10: Shorting Plug
  • M3500-opt11: K Type Thermocouple Probe

Specifications

DC CHARACTERISTICS

Function

Range

Resolution

1 Year Accuracy2

DC Voltage

100.0000 mV

0.1 µV

0.0080+0.0045

1.000000 V

1.0 µV

0.0090+0.0010

10.00000 V

10.0 µV

0.0120+0.0020

100.0000 V

100.0 µV

0.0120+0.0020

1000.000 V

1 mV

0.0130+0.0030

DCI

10.00000 mA

10 nA

0.050+0.020

100.0000 mA

100 nA

0.050+0.010

1.000000 A

1 µA

0.150+0.020

3.00000 A

10 µA

0.20+0.030

10.00000 A

10 µA

0.250+0.050

2WΩ3/4WΩ

100.0000 Ω

100 µΩ

0.020+0.005

1.000000 KΩ

1 mΩ

0.020+0.002

10.00000 KΩ

10 mΩ

0.020+0.002

100.0000 KΩ

100 mΩ

0.020+0.002

1.000000 MΩ

1 Ω

0.020+0.004

10.00000 MΩ

10 Ω

0.100+0.004

100.0000 MΩ

100 Ω

1.500+0.005

DIODE

1.00000 V

10 µV

0.020+0.020

CONTINUITY (for 2WΩ)

1000.00 Ω

10 mΩ

0.020+0.030

FREQUENCY & PERIOD

Function

Range

Frequency (Hz)

1 Year Accuracy2

Frequency & Period

100 mV to 750 V4

10-40

0.03

40–300K

0.02

AC CHARACTERISTICS

Function

Range

Resolution

Frequency (Hz)

1 Year Accuracy2

ACV (TRMS)

100.0000 mV

0.1 µV

10–20K

0.1200 +0.050

1.000000 V to 750.0000 V4

20K–50K

0.2500+0.050

50K–100K

0.6500+0.080

100K-300K

4.5000+0.500

1.0 µV to 1 mV

10–20K

0.1200+0.040

20K–50K

0.2500+0.050

50K–100K

0.6500 +0.080

100K-300K

4.5000+0.500

ACI (TRMS)

1.000000 A

1 µA

10-1K

0.20+0.04

1K-5K 1.00+0.10
3.00000 A 10 µA 10-1K 0.30 +0.060
1K-5K 1.50+0.15
10.00000 A 10 µA 10-1K 0.50+0.10
1K-5K 2.00+0.20

CAPACITANCE CHARACTERISTICS

Function

Range

Test Current

1 Year Accuracy2

CAPACITANCE5

1 nF
10 nF
100 nF
1 µF
10 µF
100 µF
1000 µF
10000 µF

10 µA
10 µA
100 µA
100 µA
100 µA
1 mA
1 mA
1 mA

2.0+0.80
1.0+0.50
1.0+0.50
1.0+0.50
1.0+0.50
1.0+0.50
1.0+0.50
2.0+0.50

TEMPERATURE CHARACTERISTICS

Function

Type

Range

1 Year Accuracy2

THERMOCOUPLE6

B
C
E
J
K
N
R
S
T

600°C~1820°C
0°C~2316°C
-250°C~1000°C
-210°C~1200°C
-200°C~1372°C
-200°C~1300°C
0°C~1767°C
0°C~1767°C
-250°C~400°C

1.5°C
1.5°C
1.5°C
1.0°C
1.0°C
1.0°C
1.5°C
1.5°C
1.5°C

General Specifications

Item

Limitation & description

Power Supply

100V/120V/220V/240V ± 10%

Power Line Frequency

50/60 Hz ± 10%

Power Consumption

25 VA peak (5 W AVERAGE)

Operating Temperature

0 °C to 50 °C

Operating Humidity

Maximum relative humidity 80% for temperature up to 31°C

Storage Temperature

- 40 °C to 70°C

Operating Altitude

Up to 2000 M

Bench Dimensions (WxHxD)

214.6x88.6x280.7mm

Weight

2.23 KG

Safety7

IEC61010-1:2001/EN61010-1:2001 (2nd Edition)
Measurement CAT II 600V, CAT I 1000V
Pollution Degree 2

EMC

EN61326-1:2006, EN61326-2-1:2006

VIBRATION

MIL-PRF-28800F, 3.8.4.2 VIBRATION, SINUSOIDAL CLASS 1,2

SHOCK

MIL-PFR-28800F, 4.5.5.4 MECHANICAL

Warranty

One Year

1The specifications are under 2-hour warm-up condition with the setting 10 PLC, and they’re relative to the calibrator specifications in PICOTEST.
2 ± (% of reading + % of range), (23 ℃ ± 5 ℃)
3 The Null function must be used when the 2WΩ is adopted.
4 The Range 750 V is limited to 100 KHz.
5 The Null function must be used.
6 The measured value must be plus the error of test leads.
7 The LO jack is marked with 500Vpk against ground and SENSE HI to LO is only marked with 200Vpk, in opposition to the label of 600V CAT II and/or 1000V CAT I against ground and IEC 61010-1.

Specifications are subject to change without notice.

All other trademarks and trade names are the property of their respective companies.

 User's Manual